| | Product Group: | Calibrators |
| Product Category: | Hand-held |
| Product Name: | Pressure calibrator |
| Product Code: | Calog-Pressure II |
| Brochure: | 1,082 Kb |
| Short Description: | Pressure and mA calibrator with serial pressure transmitters for -1 to 1000 bar ranges, 0.05% to 0.02% accuracy
|
|
|
|
Calog-Pressure II is a handheld calibrator with with one or more pressure transmitters for any range up to 1000 bar. This instrument also has the functions to install and maintain all powered and non-powered milliamp loops. It is supplied with 1 plug-on pressure sensor with optional additional sensors available for the other pressure ranges specified below.
As a small, rugged unit with rubber protective boot, high precision and Li-ion long-lasting rechargeable battery pack, Calog-Press is ideal for industrial field calibration. Data may be logged to the removable SD card in CSV format, with user-selectable data fields and timebase. Logged data may also be retrieved via the USB port. This instrument meets CE and C-Tick EMC and EMI and Low Voltage directives.
Features:
- Small, rugged, handheld with a protective rubber cover
- Graphic display of measured value, percent and battery status
- Measure, source and simulate 0-24 mA with 0.01% accuracy and 1 microamp resolution
- Measure 32V with 0.005% accuracy and 1mV resolution
- Continuity test function
- Trend logging with programmable timebase
- Displays values in mA and % or % and mA
- Measure complete vacuum to 1000 bar pressure
- Measures in bar, mbar, kPa, MPa, psi, m of H²O, in of H²O, ft of H²O, mm of Hg (torr), cm of Hg, in of Hg, kg per cm² and atmospheres
- Contains a handy pressure conversion calculator
- Programmable auto-off, restart at last setting and selectable display resolution
- Data logging to removable SD card with time and date
- USB port for access to logged data
- Lithium-ion rechargeable battery pack with 'fuel gauge' chip
- One pressure sensor included, more are optional
- Carry case, charger and test leads supplied as standard
- 2 year guarantee, excluding batteries
|
|
| |
Technical Specifications: Pressure Calibrator
|
| Digital Input Pressure Ranges |
|
|
|
-1 to 1 bar, 0 to 3 bar / 10 bar / 30 bar / 100 bar / 300 bar / 700 bar / 1000 bar with other ranges available on request
|
| Analogue Input Ranges |
Impedance |
Accuracy |
Resolution |
| 0 to 24 mA |
Input Impedance ± 1 Ω, 250mA Fused |
0.01 % |
1 µA |
| 0 to 32 V |
Input Impedance ±110 kΩ |
0.005 % |
1 mV |
Continuity with 100 Ω trigger and 1 mA excitation with visual and audible confirmation
|
| Analogue Output Ranges |
Load |
Accuracy |
Resolution |
| 0 to 24 mA |
Output load max 500 Ω |
0.01 % |
1 µA |
|
| |
|
|
|